The Symptom:
An Alaris infusion pump shows an "Incorrect Cassette" error. The nurse swears the cassette is new. The unit refuses to work full stop. 911biomed simple things go wrong work full
- What goes wrong: A nurse wiped the exterior with a bleach wipe. Bleach solution seeped through a membrane keypad.
- The Symptom: Sticky buttons or phantom key presses.
- The Full Fix: Disassemble, wash the silicone mat in distilled water, dry for 24 hours. Do not order a new keypad.
Fix the simple thing first. You will get the unit back to work full capacity faster, cheaper, and with a lot less swearing. And that is the true spirit of 911BIOMED. The 911biomed Simple Things Go Wrong Work: A